RecalledRecall 94-032Published

MJ Harris and Associates Dubby Pacifier Thermometers

M.J. Harris And Associates Recall Pacifier Thermometer

Hazard
Choking
Units
Not stated
Incidents
None listed
Remedy
See the notice

CPSC recall 94-032, announced February 17, 1994, covers MJ Harris and Associates Dubby Pacifier Thermometers. Hazard: Choking.

What to check on the product label

Compare the model on your label, then check any serial numbers, date codes, and exclusions in the notice. A model match alone does not confirm that your unit is included.

The CPSC data does not provide a discrete model-number list for this record. Compare the product description and identification details on the official notice.

Additional identifying details

The pacifier thermometers are sold in blue boxes with a sticker that reads "Made in Taiwan." The boxes have a window front and the name "Dubby" printed on the top panel.

Which products are included?

Read the full CPSC product description and restrictions

January 11, 1994 Release # 94-032 PRODUCT: "Pacifier Thermometers," brand name "Dubby." About 340 pacificer thermometers were imported from Taiwan by M.J. Harris and Associates and distributed nationwide through pharmacies and mail-order companies between April and August 1993. The pacifier thermometers are pink or blue with a clear silicon rubber nipple. They are sold in blue boxes with a window front; the name "Dubby" is printed on the top panel. PROBLEM: Because the product is both a pacifier and a thermometer, it is subject to FDA requirements and CPSC's pacifier regulations. During CPSC testing, the nipples separated from the base of the pacifier thermometers presenting a potential choking hazard to infants and young children. WHAT TO DO: Consumers should take the pacifier thermometers away from infants and young children immediately and return them to the stores or mail-order firms where purchased for a full refund. Consumers with questions about this recall may contact M.J. Harris and Associates at 1-800-531-5486 WASHINGTON, DC -- CPSC, M.J. Harris and Associates, Anacortes, WA is voluntarily recalling 340 "Pacifier Thermometers," brand named "Dubby." Because the product is both a thermometer and a pacifier, it is subject to FDA's requirements and CPSC's pacifier regulations. The pacifier thermometer fell apart during routine CPSC testing to determine compliance with CPSC's Pacifier Requirements. The pacifier's nipple separated from the base, presenting a potential choking hazard to infants and young children. The pacifiers have a built-in electronic L.C.D. thermometer that displays the temperature of the baby who is sucking on the nipple. The pacifier thermometers are blue or pink with a clear silicon rubber nipple. The pacifier thermometers are sold in blue boxes with a sticker that reads "Made in Taiwan." The boxes have a window front and the name "Dubby" printed on the top panel. A clear plastic storage cover is provided with each pacifier thermometer. The pacifier thermometers were imported by M.J. Harris and Associates from Taiwan and distributed through pharmacies and mail-order firms nationwide between April and August of this year. Each pacifier thermometer sold for as much as $12.50. Consumers are urged to take the pacifier thermometers away from babies and young children immediately and return them to the stores or mail-order firms where purchased for a full refund. Consumers with questions about this recall may contact M.J. Harris and Associates toll-free at: 1-800-531- 5486. Neither CPSC nor M.J. Harris and Associates is aware of any injuries involving these pacifiers. This voluntary recall is being conducted to prevent the possibility of injury. The potentially hazardous pacifier thermometers were identified by a competitor of M.J. Harris and Associates. CPSC is announcing this recall as part of its mission to protect the public from unreasonable risks of injury and death associated with consumer products. The Commission's objective is to reduce the estimated 28.6 million injuries and 21,700 deaths associated each year with the 15,000 different types of consumer products under CPSC's jurisdiction

Confirm these details on the official CPSC notice

Why it was recalled

Choking

Official remedy

No separate remedy text is present in the API record. Follow the official CPSC notice.

Product and sales details

Brand
MJ Harris and Associates
Manufacturer
MJ Harris and Associates, Dubby
Recall number
94-032
Manufactured in
Taiwan

Dubby Pacifier Thermometers

CPSC product type: Pacifiers & Accessories

Know someone who has one?

Recalls only work when owners hear about them. Send this notice to family or friends who might own the same product.

Related recalls