Model number check

AG-12011

Is AG-12011 recalled? Yes. Model number AG-12011 is listed in CPSC recall 27-014: Zhongshan Lebo Medical Agrish Adult Portable Bed Rails, models 12014, AG-12011 and 12016, announced October 1, 2026. Hazard: laceration, fall and entrapment. Remedy: refund.

Compare every letter, number, space, and suffix on your product label. A similar-looking model may not be included in a recall — the official CPSC notice is always the final authority.

In My homeOpen My home

Advertisements

Recall 27-014, published October 1, 2026

Zhongshan Lebo Medical Agrish Adult Portable Bed Rails, models 12014, AG-12011 and 12016

Recalled Agrish Adult Portable Bed Rail Model Number AG-12011
Recalled Agrish Adult Portable Bed Rail Model Number AG-12011Photo: CPSC recall notice

What to check on the product label

Compare the model on your label, then check any serial numbers, date codes, and exclusions in the notice. A model match alone does not confirm that your unit is included.

Affected model numbers

AG-12011
Additional identifying details

Check the product label for these CPSC-listed model numbers: AG-12011.

Which products are included?

This recall involves Agrish-branded adult portable bed rails, models AG-12011, 12014 and 12016, for use on adult beds. The model numbers are printed in the owner's manual and labeled on the bed rail's metal post. The recalled bed rails have white tubing with rounded gray foam rubber handle grips. Each model has multiple batch numbers and manufacture dates listed on the unit and the instruction manual. Recalled Bed Rails: Model Numbers Size 12014 About 11 inches wide and 39 inches in height AG-12011 12016 About 37 inches wide and 45 inches in height

Confirm these details on the official CPSC notice

Why it was recalled

The recalled bed rails violate the mandatory standard for adult portable bed rails because when the bed rails are attached to a bed, users can become entrapped within the bed rail or between the bed rail and the side of the mattress, posing a serious entrapment hazard and risk of death by asphyxiation. The bed rails do not meet structural stability or retention strap requirements, posing a fall hazard. Additionally, the bed rails' push pins and push pin holes are incorrectly sized, posing a laceration hazard.

Official remedy

Consumers should immediately stop using the recalled adult portable bed rails and contact Agrish Direct for a full refund. Consumers will be asked to go to http://www.agrish.online/recall to register for the recall and receive instructions on how to dispose of the recalled bed rail.

Remedy type: Refund

What should I do if I have AG-12011?

  1. Confirm the model number. Find the label and compare every letter and digit with AG-12011, then check the date codes and lot or batch numbers the notice lists.
  2. Stop using it. The notice says to stop using recalled units immediately.
  3. Contact the firm. Go to www.agrish.online/recall, or email agrish.us@outlook.com.
  4. Get the remedy. This recall offers a refund.
  5. Read the official notice. The CPSC notice is the final word on which units are included and what to do.
Recall published
October 1, 2026
Units
About 35,083 units (About 5,000 of these bed rails were previously recalled in January 2026)
Sold at / sales period
Sold Online At: Amazon.com from April 2025 through May 2026 for between $50 and $80.

Zhongshan Lebo Medical recall history

Zhongshan Lebo Medical is named in 2 CPSC notices in this index, all in 2026. Recall 27-014 (Oct 2026), which lists AG-12011, is the most recent of them.

One other Zhongshan Lebo Medical bed rail recall is in this index, with the model numbers it lists:

A model number missing from these lists is not proof that it is unaffected. Some notices identify products by description, date code, or UPC instead, and this index holds only CPSC notices.